Gas chromatography for determination of N-isopropylaniline in workplace atmosphere.
- Author:
Jing XUAN
1
;
Xingfu PAN
;
Xiaodong LIU
;
Huifang YAN
2
;
Yanshu ZHANG
3
Author Information
- Publication Type:Journal Article
- MeSH: Air; analysis; Air Pollutants, Occupational; analysis; Aniline Compounds; analysis; Chromatography, Gas; methods; Workplace
- From: Chinese Journal of Industrial Hygiene and Occupational Diseases 2015;33(2):143-145
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo establish a method for determination of N-isopropylaniline in the workplace atmosphere by gas chromatography.
METHODSAir samples were collected by silica gel tube and desorbed by acetone. Then they were separated through DB-WAX columns and N-isopropylaniline was determined by flame ionization detector.
RESULTSThe concentration of N-isopropylaniline showed a good linear relationship within the range of 1.40∼665.0 µg/ml. The sampling efficiency was 100%. The accuracy was 96%∼ 99% and the precision was 2.1%∼7.0%. The minimum detectable concentration was 0.056 mg/m(3) (with sampled air volume of 7.5 L).
CONCLUSIONThe method meets the requirements of analysis and applies to the determination of N-isopropylaniline in the workplace atmosphere.