Study on the activity--heart rate prediction system for motion sensed rate responsive pacemakers.
- Author:
Ping GUO
1
;
Weixin SUN
;
Jie JIN
;
Qingping ZHAO
;
Xiang CHEN
;
Shu KONG
;
Yizhuo HUANG
Author Information
1. Institute of Medical Electronics in Medical School, Key Laboratory of Biomedical Information Engineering, Ministry of Education, Xi' an Jiaotong University, Xi' an, Shaanxi, China. guoping6687@mail.xjtu.edu.cn
- Publication Type:Journal Article
- MeSH:
Algorithms;
Cardiac Pacing, Artificial;
methods;
Heart Rate;
physiology;
Humans;
Motor Activity;
Pacemaker, Artificial;
Signal Processing, Computer-Assisted
- From:
Chinese Journal of Medical Instrumentation
2011;35(5):324-329
- CountryChina
- Language:Chinese
-
Abstract:
Focusing on the heart rate control mode of acceleration sensor based rate responsive pacemakers, this paper implemented the design of activity--heart rate prediction system. Bluetooth module was used as communication means in activity--heart rate prediction system, and the slave computer was used to complete the acceleration signal acquisition and processing, map from acceleration signal to the pacing rate signal, and achieve real-time transmission of acceleration signal and heart rate signal. The master computer fulfilled real-time display and recording of acceleration signal and heart rate signal, moreover, it achieved control function to the slave computer algorithm through classification of 6 parameters. The results of verification experiment showed that there was a significant relation between mapping heart rate and actual heart rate using linear mapping algorithm (R2 = 0.787, P < 0.001).