Investigation of Elekta linac characteristics for VMAT.
- Author:
Guangwen LUO
1
;
Kunyi ZHANG
Author Information
- Publication Type:Journal Article
- MeSH: Humans; Particle Accelerators; Radiotherapy Dosage; Radiotherapy, Intensity-Modulated; instrumentation; methods
- From: Chinese Journal of Medical Instrumentation 2012;36(1):22-24
- CountryChina
- Language:Chinese
-
Abstract:
UNLABELLEDThe aim of this study is to investigate the characteristics of Elekta delivery system for volumetric modulated arc therapy (VMAT). Five VMAT plans were delivered in service mode and dose rates, and speed of gantry and MLC leaves were analyzed by log files. Results showed that dose rates varied between 6 dose rates. Gantry and MLC leaf speed dynamically varied during delivery.
CONCLUSIONThe technique of VMAT requires linac to dynamically control more parameters, and these key dynamic variables during VMAT delivery can be checked by log files. Quality assurance procedure should be carried out for VMAT related parameter.