Evaluation of a portable device based on peripheral arterial tone in the detection of obstructive sleep apnea
10.3760/cma.j.issn.1673-0860.2012.02.006
- VernacularTitle:腕式睡眠监测仪在阻塞性睡眠呼吸暂停低通气综合征诊断中的应用
- Author:
Juan ZHANG
1
;
Chun-Yan WANG
;
Ning-Yu WANG
;
Jun-Qing LIU
;
Jing-Yan DU
;
Guo-Xia ZHOU
;
Zhan-Feng YAN
;
Xiao-Hui WEN
;
Li LI
;
Jin-Feng LIU
Author Information
1. 首都医科大学附属北京朝阳医院
- Keywords:
Sleep apnea,obstructive;
Polysomnography;
Monitoring,ambulatory
- From:
Chinese Journal of Otorhinolaryngology Head and Neck Surgery
2012;47(2):112-116
- CountryChina
- Language:Chinese
-
Abstract:
Objective To assess the accuracy of a wrist-worn device (Watch-PAT 200) in the diagnosis of obstructive sleep apnea syndrome (OSAHS). Methods Forty-three adult subjects with suspected OSAHS simultaneously had a standard in-laboratory polysomnogram (PSG) and wore the WatchPAT 200 during a full-night recording.PSG sleep and respiratory events were scored according to standard criteria. The PSG recordings were blindly manually analyzed, while Watch-PAT data were scored automatically based on the algorithm developed previously.Results The mean age of the subjects was (42.2 ± 12.2) years ((x) ±s),and mean body mass index was (28.0 ±3.9) kg/m2.Mean PSG apnea hypopnea index (AHI) was (34.9 ± 29.9) events per hour,and mean PAT-AHI was (36.0 ± 29.2) events per hour.There was a significant correlation between PAT AHI and AHI by PSG (r =0.931,P <0.01 ).A Bland-Altman plot of PAT AHI and PSG AHI was also used to assess the accuracy of Watch-PAT 200.At lower levels of AHI,PAT tended to overestimate disease severity,while at higher levels of AHI,Watch-PAT underestimated severity.To assess sensitivity and specificity of Watch-PAT,constructed receiver operator characteristic curves using a variety of AHI threshold values (5,15 and 30 events per hour).For AHI ≥5events per hour as threshold value,the Watch-PAT diagnosing rate was 93%,and sensitivity as well as specificity were 94.7% and 80.0%.The misdiagnosis rate and missed diagnosis rate were 20.0% and 5.3%.Optimal combinations of sensitivity and specificity for the AHI threshold values ( 15 and 30 events per hour) were 82.6% and 100.0%,95.0% and 95.7% respectively.Conclusion The Watch-PAT 200 may offer an accurate,robust,and reliable ambulatory method for the detection of OSAHS,with minimal patient discomfort.