Detection of surface EMG signal using active electrode.
- Author:
Qinghua HE
1
;
Chenglin PENG
;
Baoming WU
;
He WANG
Author Information
1. 5th Department of Research Institute of Surgery, Daping Hospital, Third Military Medical University, Bioengineering College, Chongqing University, Chongqing.
- Publication Type:Journal Article
- MeSH:
Electrodes;
Electromyography;
instrumentation;
methods;
Equipment Design;
Signal Processing, Computer-Assisted;
instrumentation
- From:
Journal of Biomedical Engineering
2003;20(3):488-490
- CountryChina
- Language:Chinese
-
Abstract:
Research of surface electromyogram(EMG) signal is important in rehabilitation medicine, sport medicine and clinical diagnosis, accurate detection of signal is the base of quantitative analysis of surface EMG signal. In this article were discussed how to reduce possible noise in the detection of surface EMG. Considerations on the design of electrode unit were presented. Instrumentation amplifier AD620 was employed to design a bipolar active electrode for use in surface EMG detection. The experiments showed that active electrode could be used to improve signal/noise ratio, reduce noise and detect surface EMG signal effectively.