Measurement of Serum Levels of 25-Hydroxyvitamin D3 and 25-Hydroxyvitamin D2 Using Diels-Alder Derivatization and Ultra-Performance Liquid Chromatography-Tandem Mass Spectrometry.
- Author:
Hyungsuk KIM
1
;
Sun Hee JUN
;
Taeksoo KIM
;
Sang Hoon SONG
;
Kyoung Un PARK
;
Junghan SONG
Author Information
- Publication Type:Original Article
- Keywords: Vitamin D; 4-phenyl-1,2,4-triazoline-3,5-dione (PTAD); Tandem mass spectrometry; Ultra performance liquid chromatography; Immunoassay
- MeSH: 25-Hydroxyvitamin D 2; Acetates; Calcifediol; Calibration; Formates; Humans; Immunoassay; Liquid-Liquid Extraction; Mass Spectrometry; Quaternary Ammonium Compounds; Tandem Mass Spectrometry; Triazoles; Vitamin D
- From:Laboratory Medicine Online 2012;2(4):188-196
- CountryRepublic of Korea
- Language:Korean
- Abstract: BACKGROUND: Vitamin D has been recently shown to play important roles in the functioning of various systems. Most of the current analytical methods for measuring vitamin D levels are based on immunoassays. We simultaneously measured the levels of 25-hydroxyvitamin D3 [ 25(OH)D3 ] and 25-hydroxyvitamin D2 [ 25(OH)D2 ] in human serum by performing ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) after Diels-Alder derivatization with 4-phenyl-1,2,4-triazoline-3,5-dione (PTAD) and evaluated the performance of our method. METHODS: After liquid-liquid extraction, samples were dried under N2 at 50degrees C for 1 hr followed by Diels-Alder derivatization with ethyl acetate containing 0.1 mg/mL PTAD. The samples were resuspended in 60 microL of methanol:10 mM ammonium formate solution (1:1, V/V). C18 UPLC column and positive ion multiple reaction monitoring transitions such as m/z 558.35-->298.1, 25(OH)D3; m/z 570.35-->298.1, 25(OH)D2; and m/z 564.35-->298.1, hexadeuterated-25(OH)D3 were used for UPLC-MS/MS. RESULTS: The within-run imprecision (CVs) for 25(OH)D3 and 25(OH)D2 were 3.5-4.0% and 3.8-4.2%, respectively, and the corresponding between-run CVs were 3.3-5.5% and 4.7-5.8%. The lower limit of quantification for 25(OH)D3 and 25(OH)D2 were 0.5 and 1.0 ng/mL, respectively. The curve for interassay calibration variability data obtained over concentrations of 0-120 ng/mL for 25(OH)D3 and 0-90 ng/mL for 25(OH)D2 was linear and reproducible [ 25(OH)D3, R2=0.993; 25(OH)D2, R2=0.998]. The total 25(OH)D levels in Koreans (average, 18.7 ng/mL) were lower than those in American Caucasians, and the percentage of people with total 25(OH)D levels under 10 ng/mL was 8.1%. CONCLUSIONS: Our method to measure 25(OH)D3 and 25(OH)D2 levels by performing UPLC-MS/MS after PTAD derivatization showed good performance as a sensitive and reproducible method for routine analysis of vitamin D status.