Investigation on Mobile Phone Based Thermal Imaging System and Its Preliminary Application.
- Author:
Fufeng LI
;
Feng CHEN
;
Jing LIU
- Publication Type:Journal Article
- MeSH:
Algorithms;
Body Temperature;
Cell Phone;
Humans;
Infrared Rays;
Mobile Applications
- From:
Chinese Journal of Medical Instrumentation
2015;39(3):173-176
- CountryChina
- Language:Chinese
-
Abstract:
The technical structure of a low-cost thermal imaging system (TIM) lunched on a mobile phone was investigated, which consists of a thermal infrared module and mobile phone and application software. The designing strategies and technical factors toward realizing various TIM array performances are interpreted, including sensor cost and Noise Equivalent Temperature Difference (NETD). In the software algorithm, a mechanism for scene-change detection was implemented to optimize the efficiency of non-uniformity correction (NUC). The performance experiments and analysis indicate that the NETD of the system can be smaller than 150 mK when the integration time is larger than 16 frames. Furthermore, a practical application for human temperature monitoring during physical exercise is proposed and interpreted. The measurement results support the feasibility and facility of the system in the medical application.