Sevoflurane Output in the Isoflurane/Halothane Diamedica Draw-over Vaporiser.
- Author:
Woo Chiao TAY
1
;
Wah Tze TAN
Author Information
- Publication Type:Journal Article
- MeSH: Anesthesia, Inhalation; instrumentation; methods; Anesthetics, Inhalation; administration & dosage; Halothane; Humans; Isoflurane; Methyl Ethers; administration & dosage; Nebulizers and Vaporizers; Temperature
- From:Annals of the Academy of Medicine, Singapore 2015;44(7):252-256
- CountrySingapore
- Language:English
-
Abstract:
INTRODUCTIONDraw-over anaesthesia remains an attractive option for conduct of anaesthesia in austere conditions. The Diamedica Draw-over Vaporiser (DDV) is a modern draw-over vaporiser and has separate models for isoflurane/halothane and sevoflurane.
MATERIALS AND METHODSA laboratory study was done to measure sevoflurane output in an isoflurane/ halothane DDV. We did 3 series of experiments with the isoflurane/halothane DDV. We measured anaesthetic agent output in both push-over and draw-over setups, and at minute ventilation of 6 L/min and 3 L/min. Series 1 experiment was done with isoflurane in the DDV at ambient temperature of 20°C. Series 2 experiment was done with sevoflurane in the DDV at ambient temperature of 20°C. Series 3 experiment was done with sevoflurane in the DDV and with the DDV placed in a water bath of 40°C.
RESULTSThe sevoflurane output was found to be two-thirds of the isoflurane/ halothane DDV dial setting at ambient temperature of 20°C. With the DDV in a 40°C water bath, the sevoflurane output was found to be about the isoflurane/ halothane DDV dial settings.
CONCLUSIONIn our experiment, we show that it is possible to use sevoflurane in an isoflurane/halothane DDV.