Comparison of two different methods to detect HIV antibodies
10.3760/cma.j.issn.1003-9279.2012.06.028
- VernacularTitle:两种不同方法检测HIV抗体结果的比较
- Author:
Jing-Xia GUO
1
;
Jun XU
;
Lin CHEN
;
Jia LIU
;
Jing ZHAO
;
Yong-Ji SONG
;
Ai-Xia LIU
;
Bo-An LI
;
Yuan-Li MAO
Author Information
1. 解放军第三○二医院临床检验中心
- Keywords:
HIV;
Enzyme-linked immunosorbent assay;
Chemiluminescence method
- From:
Chinese Journal of Experimental and Clinical Virology
2012;26(6):492-493
- CountryChina
- Language:Chinese
-
Abstract:
Objective Evaluated the chemiluminescence and enzyme-linked immunosorbent assay (ELISA) to detect HIV antibodies,and compared the results,to provide a reference for the selection and clinical application of HIV screening.Methods 3000 cases of our hospital patients were measured by enzyme-linked immunosorbent assay and chemiluminescence immunoassay,using comfirmming experimental results as gold standards.Comparing sensitivity,specificity and other Indicators.Results In the diagnosis of HIV infection,enzyme-linked immunosorbent assay and chemiluminescence immunoassay had no significant difference.The positive rate of enzyme-linked immunosorbent assay was 0.93%,while the sensitivity and specificity were 89.66%,99.93%,the positive rate of chemiluminescence immunoassay was 1.03%,while the sensitivity and specificity were 100%,99.93%,respectively.Conclusion Both methods are suitable for screening of HIV,having high specificity,and chemiluminescence has greater sensitivity than ELISA.