Comparison of statics characteristics between anterior transpedicular screws system and vertebral body screws system in lower cervical spine.
- Author:
Liu-Jun ZHAO
;
Jie LI
;
Wei-Yu JIANG
;
Rong-Ming XU
;
Wei-Hu MA
;
Feng QI
;
Liang YU
;
Ming ZHANG
;
Yi-Jin WANG
- Publication Type:Journal Article
- MeSH: Biomechanical Phenomena; Bone Plates; Bone Screws; Cervical Vertebrae; surgery; Humans; Internal Fixators
- From: China Journal of Orthopaedics and Traumatology 2014;27(2):118-122
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo compare statics characteristics between anterior transpedicular screws (ATPS) system and vertebral body screws (VBS) system in lower cervical spine.
METHODSSixteen fresh cervical specimens were collected and dissected into 32 different units (functional spinal unit, FSU), 8 units in C3,4, C4,5, C5,6 and C6,7 each. The subjects were randomly divided into group A and B. The anterior transpedicular screw-plate system and anterior vertebral body screw-plate system were implanted separately in group A and B. Then, the maximum axial pull out strength was tested and compared between two fixation system.
RESULTSMaximum pull out strength was (604.68 +/- 48.76) N in group A and (488.24 +/- 32.42) N in group B, and there was significant difference between two groups (t = 2.147, P < 0.05). There was no statistically significant difference in all FSU between anterior transpedicular screws system and vertebral body screws system (F(A) = 2.27, F(B) = 2.05, P > 0.05).
CONCLUSIONThe pull out strength of anterior transpedicular screws system is better than vertebral body screws system, and the anterior transpedicular screws has the biomechanical feasibility in clinic.