Dosimetry by Using EBT2 Film for Total Skin Electron Beam Therapy (TSET).
- Author:
Ui Jung HWANG
1
;
Jeong Eun RAH
;
Hojin JEONG
;
Sung Hwan AHN
;
Dong Wook KIM
;
Sang Yeob LEE
;
Young Gyung LIM
;
Myonggeun YOON
;
Dong Ho SHIN
;
Se Byeong LEE
;
Sung Young PARK
;
Hong Ryull PYO
;
Weon Kuu CHUNG
Author Information
1. Proton Therapy Center, National Cancer Center, Seoul, Korea.
- Publication Type:Original Article
- Keywords:
TSET;
EBT2 film;
TLD;
Dosimeter;
In-vivo
- MeSH:
Electrons;
Humans;
Particle Accelerators;
Skin
- From:Korean Journal of Medical Physics
2010;21(1):60-69
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
For treatment of Total Skin Electron beam Therapy (TSET), measurement of dose at various conditions is need on the contrary to usual radiotherapy. When treating TSET with modified Stanford technique based on linear accelerator, the energy of treatment electron beam, the spatial dose distribution and the actual doses deposited on the surface of the patient were measured by using EBT2. The measured energy of the electron beam was agreed with the value that measured by ionization chamber, and the spatial dose distribution at the patient position and the doses at several point on the patient's skin could be easily measured by EBT2 film. The dose on the patient that was measured by EBT2 film showed good agreement with the data measured simultaneously by TLD. With the results of this study, it was proven that the EBT2 film can be one of the useful dosimeter for TSET.