A retrospective study of six fractured platform-switching abutments
10.3760/cma.j.issn.1002-0098.2015.09.004
- VernacularTitle:平台转换种植修复基台折断六例原因分析
- Author:
Yibo ZHU
1
;
Lixin QIU
;
Jianhui LI
;
Ye LIN
Author Information
1. 100025,北京大学口腔医学院·口腔医院第四门诊部
- Keywords:
Dental implantation;
Platform-switching;
Abutment fracture
- From:
Chinese Journal of Stomatology
2015;50(9):527-530
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the reasons for platform-switching abutment fracture,the treatment and prevention methods.Methods Between March 2009 and December 2014,all of the fractured platform-switching abutments were evaluated.The fracture rate of platform-switching abutments was calculated and the gender,age,implant site,implant size,abutment size,prosthetic type,loading time and crown materials were analyzed.Results A total of 2 487 platform-switching implants were placed during this period and 6 abutments were fractured.The fracture rate was 0.24% (6/2 487).All of the 6 fractured abutments were placed in molar region.All occlusal planes were made of metal material.Five patients were male.All the fracture abutments could not be removed and the implants had to be removed and reimplanted.Conclusions Without special abutment removing instruments,it is difficult to remove the fractured platform-switching abutment.In molar region,platform-switching implant system should be carefully used.Increasing implants number and splinting crowns with porcelain occlusal planes might reduce the abutment fracture risk.