Research on the feasibility of a magnetic-coupling-driven axial flow blood pump.
- Author:
Xiaoqing YU
1
;
Wenxiang DING
;
Wei WANG
;
En CHEN
;
Zuming JIANG
;
Wenyan ZOU
Author Information
1. Department of Pediatric Thoracic Cardiovascular Surgery, Shanghai Children's Medical Center, Xinhua Hospital, Shanghai Second Medical University, Shanghai 200127, China.
- Publication Type:Journal Article
- MeSH:
Assisted Circulation;
instrumentation;
Equipment Design;
Extracorporeal Circulation;
instrumentation;
Hydrodynamics;
Magnetics
- From:
Journal of Biomedical Engineering
2004;21(1):131-133
- CountryChina
- Language:Chinese
-
Abstract:
A new-designed axial flow blood pump, dived by magnetic coupling and using internal hollow brushless DC motor and inlet and outlet in line with impeller, was tested in mimic circuit. The results showed good performance of the new pump and indicated that its hydrodynamic characteristic can meet the demands of clinical extracorporeal circulation and auxiliary circulation.