Effects on performance of high-speed dental handpieces subjected to autoclaving.
- Author:
Xiao-hong DENG
1
;
Zheng SUN
;
Jing SU
;
Hong QIAO
;
Xiao XIAO
Author Information
- Publication Type:Journal Article
- MeSH: Dental High-Speed Equipment; Equipment Contamination; prevention & control; Sterilization; methods
- From: Chinese Journal of Preventive Medicine 2006;40(4):285-289
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo investigate the effects on performances of high-speed dental handpieces subjected to autoclaving.
METHODSThe method of trial used to evaluate the effect of autoclaving included that ten chemical and ten biological indicators were arranged in each trial group being set by one sort of temperature and sterilization time. The range of temperature was from 100 degrees C to 134 degrees C and the sterilization time was from 1 to 45 minutes. The reliability of autoclaving was evaluated by the result of chemical indicator and the endospores of Bacillus stearothermophilus subjected to autoclaving. Ten newly imported and ten new domestic dental handpieces in two different brands were used to carry out this trial. The effects produced by autoclaving on performances of dental handpieces subjected to different type autoclaves were evaluated by some parameters.
RESULTSThe eligible outcome in killing test of the 221 sterilized dental handpieces proved that the sterilization effect of autoclaving on dental handpieces was reliable. The chemical indicators showed that under the condition of 121 degrees C it needed 20 minutes to reach sterilization standard in the inner package and only 15 minutes on the outer package, and under the condition of 134 degrees C it took 4 minutes to reach sterilization standard in the inner package while only 2 minutes on the outer package. This outcome proved that material with package was more time-consuming than that without package during autoclaving. Autoclaving has positive effects on the performance of dental handpieces, while the autoclave with less-time full sterilization cycle has less effect on it. The performance of dental handpieces sterilized by the autoclave with 6-minute full sterilization cycle might be restorable within certain times of sterilization cycle.
CONCLUSIONThe sterilization effect of autoclaving is reliable. In order to reduce the influence on the performance of dental handpieces subjected to autoclaving, it should be reasonable to use the autoclave by which a full sterilization cycle might take less time and no use of package.