Method for determining 2, 4-D butylate in serum by gas chromatography.
- Author:
Baoying JIN
1
;
Yi QIAN
1
;
Shuming DU
1
Author Information
- Publication Type:Journal Article
- MeSH: Chromatography, Gas; methods; Humans; Serum; chemistry; Thiocarbamates; blood
- From: Chinese Journal of Industrial Hygiene and Occupational Diseases 2014;32(2):145-146
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo establish a method for determining 2, 4-D butylate in serum by gas chromatography (GC)and to provide a basis for the diagnosis and treatment of clinical poisoning.
METHODSSerum 2, 4-D butylate level was determined by the following steps: mixing serum (0.5 ml)with trichloromethane (2.0 ml), adequately shaking for extraction, standing for 5 min, centrifuging at 4 000 rpm for 10 min, blow-drying the trichloromethane layer with nitrogen, adding ethanol (50 µl)to a certain volume, adding the sample (1.0 µl), and performing GC with a hydrogen flame ionization detector.
RESULTSSerum 2, 4-D butylate level showed a linear relationship within 5∼40 µg/ml, with a regression equation of y = 1 831.6.4x-899.4 (r = 0.999 2); the minimum detectable concentration was 1.0 µg/ml. The recovery rate was 88.7%∼103.0% (relative standard deviation (RSD) 3.8%∼5.0%). The intra-day RSD and inter-day RSD were 3.87-4.92% and 3.33%∼5.34%, respectively.
CONCLUSIONThis determination method is simple, efficient, and accurate and provides a good means for rapid diagnosis and treatment of 2, 4-D butylate poisoning.