Application of 70 kV Third-generation High-pitch Dual-source Coronary CT Angiography in Patients with Different Body Mass Index.
- VernacularTitle:第3代双源CT 70 kV大螺距扫描冠状动脉CT血管成像在不同体质量指数患者中的应用
- Author:
Yan YI
1
;
Jian CAO
1
;
Lu LIN
1
;
Lingyan KONG
1
;
Shu JIANG
1
;
Xiao LI
1
;
Peijun LIU
1
;
Ming WANG
1
;
Man WANG
1
;
Yun WANG
1
;
Zhengyu JIN
1
;
Yining WANG
1
Author Information
- Publication Type:Journal Article
- MeSH: Body Mass Index; China; Computed Tomography Angiography; methods; Contrast Media; Heart Rate; Humans; Prospective Studies; Radiation Dosage; Radiographic Image Interpretation, Computer-Assisted; Signal-To-Noise Ratio
- From: Acta Academiae Medicinae Sinicae 2017;39(1):42-48
- CountryChina
- Language:English
-
Abstract:
Objective To investigate the optimized range of body mass index (BMI) selection for patients undergoing 70 kV high-pitch dual-source coronary CT angiography (CCTA) on the third-generation dual-source CT (DSCT). Methods Patients undergoing prospective high-pitch ultra-low contrast media (CM) CCTA on the third-generation DSCT using the automatic tube voltage selection at 70 kV were included and divided into three groups:group A,with BMI≤24 kg/m;group B,with 24 kg/m
0.05). Subjective image quality in group A(Z=2.91,P=0.004) and B(Z=2.27,P=0.021) were both significantly better than that in group C. Conclusion The ultra-low tube voltage (70 kV) combined with ultra-low CM CCTA protocol on third-generation high-pitch DSCT may be better for patients with BMI<28 kg/m2 than those with BMI ≥28 kg/min China.