Determination of trimethyltin chloride in urine by headspace-gas chromatography.
- Author:
Zheng RUAN
1
;
Hong-fang TANG
;
Dan-hua LIU
;
Cheng-min XU
;
Ya-Ling QIAN
Author Information
- Publication Type:Journal Article
- MeSH: Chromatography, Gas; methods; Humans; Trimethyltin Compounds; urine; Urinalysis; methods
- From: Chinese Journal of Industrial Hygiene and Occupational Diseases 2011;29(2):141-144
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo establish a detection method for trimethyltin chloride in urine by the Head space-GC.
METHODAfter derivatizing trimethyltin chloride, the urines was separated by the head space-gc, and then the trimethyltin chloride detected qualitatively and quantificationally.
RESULTSIn the concentration range of 0.02 ∼ 0.40 mg/L urinary trimethyltin chloride, showed a quadratic, r = 0.9992, detection limit was 0.005 mg/L, the relative standard deviation was 1.9% ∼ 2.5%, recovery was 92.0% to 100%, the urine samples can be saved at least 90 days in -18°C refrigerator.
CONCLUSIONThe instrument, reagents involved in the detection require low, the operations to processing samples are simple, high sensitivity, less interference, good reproducibility, and suitable for quantitative and qualitative analysis, convenient to promotion.