Comparison of cryoablation and radiofrequency ablation for treating atrioventricular nodal reentrant tachycardia
10.3760/cma.j.issn.0253-3758.2011.07.005
- VernacularTitle:房室结折返性心动过速冷冻消融与射频消融治疗对比分析
- Author:
Ya-Hui DING
1
;
Bai-Ming QU
;
Xian-Da CHE
;
Jian-Wu YU
;
Chang-Hua WANG
;
Hui WANG
Author Information
1. 浙江省人民医院
- Keywords:
Tachycardia,atrioventricular nodal reentry;
Catheter ablation
- From:
Chinese Journal of Cardiology
2011;39(7):625-627
- CountryChina
- Language:Chinese
-
Abstract:
Objective To compare the efficacy and safety between cryoablation (Cryo) and radiofrequency (RF) ablation for treating patients with atrioventricular nodal reentrant tachycardia (AVNRT). Methods Patients with AVNRT (n=304) were divided into Cryo group (n=67) and RF group (n=237). The procedure success rate, complete slow pathway block rate, atrioventricular block rate and relapse rate were compared between two groups. Results There was no statistically difference between 2 groups in the success rate (Cryo group 98.5% vs RF group 97.0%, P=0.820), complete slow pathway block rate (Cryo group 98.5% vs RF group 91.6%, P=0.088), atrioventricular block rate (Cryo group 0 vs RF group 2.5%, P=0.413), relapse rate (Cryo group 0 vs RF group 1.7%, P=0.643). But Cryo group had more advantage than RF group. Conclusion Efficacy and safety were comparable between cryoablation and radiofrequency ablation for treating patients with AVNRT.