Studies on the mechanical properties of thermoplastic polymer resin materials for medical outside fixation.
- Author:
Jie CHEN
1
;
Shubao ZHANG
;
Xu CHEN
Author Information
1. Department of Radiotherapy, Tianjin Cancer Hospital and Institute, Tianjin 300060, China. tjcjvip@126.com
- Publication Type:Journal Article
- MeSH:
External Fixators;
Materials Testing;
Mechanical Phenomena;
Polymers;
chemistry;
Resins, Synthetic;
chemistry;
Temperature;
Tensile Strength
- From:
Journal of Biomedical Engineering
2012;29(3):470-473
- CountryChina
- Language:Chinese
-
Abstract:
Temperature scanning tests and uniaxial tensile tests at room temperature under different strain rates for three kinds of existing low-temperature outside fixed thermoplastic medical polymer resin materials (A, B and C) were conducted on dynamic mechanical analyzer and micro material testing machine to contrast their mechanical properties. It is shown that the glass transition temperatures of these three materials are all around 65 degrees C, but material C has the widest range of shaping temperature. Stress-strain responses of the three materials are rate-dependent and material C shows better plasticity.