Effect of dialysis time in vivo on recovery of amino acids for micro-dialysis probe.
- Author:
Ke-ping ZHANG
1
;
Heng-yi ZHANG
;
Xiao-xiang ZHENG
Author Information
- Publication Type:Journal Article
- MeSH: Amino Acids; analysis; Animals; Aspartic Acid; analysis; Brain Chemistry; Chromatography, High Pressure Liquid; methods; Dialysis Solutions; analysis; Glutamic Acid; analysis; Microdialysis; methods; Rats; Rats, Sprague-Dawley; Time Factors; gamma-Aminobutyric Acid; analysis
- From: Journal of Zhejiang University. Medical sciences 2006;35(6):642-647
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo investigate the variety of vitro recovery of amino acids for microdialysis probe after different dialysis time in vivo.
METHODSProbes were dialyzed in the amino acids standard solutions with microdialysis system,amino acid standard solutions and the microdialysate of probe were detected by the method of precolumn derivation with HPLC-RF.
RESULTAfter using different time of probe made by regenerated cellulose membrane, the vitro recoveries of Asp, Glu and GABA were not completely same (Asp: F=19.669, P=0.000; Glu: F=103.955, P=0.000; GABA: F=3.454, P=0.040); while the vitro recovery of Tau had no obvious difference(F=2.001, P=0.152). After using 6 h in vivo, recovery remain percentage (RRP) of Asp, Glu,Tau and GABA was 64.34 %, 67.36%, 103.11 % and 98.23 %, respectively, the recoveries of Asp, Glu decreased obviously (Asp: P < 0.01,Glu: P <0.05). After using 12 h in vivo, the RRP of Asp, Glu, Tau and GABA was 43.44 %, 24.42%, 77.45 % and 67.36 %, respectively, the recoveries of Asp, Glu and GABA decreased obviously (Asp: P < 0.001, Glu: P < 0.001, GABA: P < 0.05). After using 24 h in vivo, the RRP of Asp, Glu,Tau and GABA was 36.26 %, 12.24 %, 89.48 % and 71.35 %, respectively, the recoveries of Asp, Glu, GABA decreased obviously (Asp: P < 0.0001, Glu: P < 0.0001, GABA: P < 0.01).
CONCLUSIONDialysis in vivo could lead to the decline of recovery of probe, the decline is more obvious after longer dialysis. So when making brain dialysis experiments, the use time of probe should not be too long. To improve the validity of data, some calibration should be made on the recoveries of probe.