Optimization of atrioventricular delay by surface electrocardiography during dual chamber pacing.
- Author:
Li-qun WU
1
;
Gang GU
;
Min CAO
;
Yong-chu SHEN
;
Kan SU
;
Wei-feng SHEN
Author Information
- Publication Type:Journal Article
- MeSH: Adult; Aged; Aged, 80 and over; Atrioventricular Node; physiopathology; Cardiac Pacing, Artificial; methods; Electrocardiography; methods; Female; Heart Block; physiopathology; therapy; Humans; Male; Middle Aged; Time Factors
- From: Chinese Medical Journal 2006;119(6):454-457
- CountryChina
- Language:English
-
Abstract:
BACKGROUNDNominal atrioventricular (AV) interval in dual chamber pacemaker (DDD) is not the best AV delay in the majority of patients with atrioventricular block. To find a simple method for optimizing AV delay adjustment, we assessed surface electrocardiography (ECG) for optimizing AV delay during dual chamber pacing.
METHODSDDD pacemakers were implanted in 46 patients with complete, or almost complete, AV block. Optimal AV delay was achieved by programming an additional delay of 100 ms, to the width of intrinsic P wave or to the interval between pacing spike to the end of P wave on surface ECG. Left ventricular (LV) end diastolic and end systolic volumes, ejection fraction and diastolic parameters were measured by Doppler echocardiography during both nominal and optimal AV delay pacing.
RESULTSCompared to nominal AV delay setting, LV end diastolic volume increased [to (53.2 +/- 11.3) ml from (50.2 +/- 10.2) ml, P < 0.05], end systolic volume decreased [to (26.1 +/- 9.0) ml from (27.9 +/- 8.2) ml, P < 0.05] during adjusted AV delay pacing, resulting in an increase in LV ejection fraction [to (68.2 +/- 5.3)% from (64.5 +/- 4.3)%, P < 0.05]. LV diastolic filling and isovolumic relaxation time were not significantly changed.
CONCLUSIONOptimization of AV delay by surface ECG is a simple method to improve LV systolic function during dual chamber pacing.