Research on Constant-current Characteristics of Howland Current Source Used in Bioelectrical Impedance Detection.
- Author:
Xingjian LIN
;
Weijie ZHAO
;
Xiaojuan LIU
;
Lihua LI
- Publication Type:Journal Article
- MeSH:
Electric Impedance;
Electric Power Supplies;
Electricity
- From:
Journal of Biomedical Engineering
2015;32(2):290-294
- CountryChina
- Language:Chinese
-
Abstract:
This study aims to analyze and improve Howland current source circuit and to study the constant current source of alternating current with high output impedance and high stability. A simulation study was carried out on the constant-current characteristics of Howland current source from two aspects of resistance match value and the selection of Op amp parameters, and then the output impedance was analyzed. The simulation experiment showed that when it was with the best matched resistance, the constant-current characteristics of Howland current source was better. Op amp parameters could affect the constant-current characteristics of Howland current source. In Howland current source circuit, the current source after parameter optimization had better output impedance and load capacity. The results showed that there was the best matched resistance in the selection by Howland current source. The Op amp with large broadband, high slew rate and open loop gain, and wide range of power supply voltage is more suitable for the bioelectrical impedance detection circuit.