Measurement Error Analysis and Calibration Technique of NTC - Based Body Temperature Sensor.
- Author:
Chi DENG
;
Wei HU
;
Shengxi DIAO
;
Fujiang LIN
;
Dahong QIAN
- Publication Type:Journal Article
- MeSH:
Body Temperature;
Calibration;
Humans;
Monitoring, Physiologic;
methods
- From:
Chinese Journal of Medical Instrumentation
2015;39(6):395-399
- CountryChina
- Language:Chinese
-
Abstract:
A NTC thermistor-based wearable body temperature sensor was designed. This paper described the design principles and realization method of the NTC-based body temperature sensor. In this paper the temperature measurement error sources of the body temperature sensor were analyzed in detail. The automatic measurement and calibration method of ADC error was given. The results showed that the measurement accuracy of calibrated body temperature sensor is better than ± 0.04 degrees C. The temperature sensor has high accuracy, small size and low power consumption advantages.