Design of a pulse oximeter used to low perfusion and low oxygen saturation.
- Author:
Shuangping TAN
1
;
Zhiguang AI
;
Yuxing YANG
;
Qingguo XIE
Author Information
1. School of Life Science & Technology, Huazhong University of Science & Technology, Wuhan 430074. tanshuangping@gmail.com
- Publication Type:Journal Article
- MeSH:
Artifacts;
Equipment Design;
Oximetry;
instrumentation;
Software
- From:
Chinese Journal of Medical Instrumentation
2013;37(3):189-196
- CountryChina
- Language:Chinese
-
Abstract:
This paper presents a new pulse oximeter used to low perfusion at 0.125% and wide oxygen saturation range from 35% to 100%. In order to acquire the best PPG signals, the variable gain amplifier(VGA) is adopted in hardware. The self-developed auto-correlation modeling method is adopted in software and it can extract pulse wave from low perfusion signals and remove motion artifacts partly.