In vitro cytotoxicity test of medical ultrasonic couplant.
- Author:
Jingxian GAO
1
;
Shasha WANG
;
Meng JIN
;
Xiaoli YAN
Author Information
1. Jiangsu Province Medical Instrument Testing Institute, Nanjing 210022.
- Publication Type:Journal Article
- MeSH:
Animals;
Cell Line;
Colorimetry;
Cytotoxicity Tests, Immunologic;
methods;
Mice;
Ultrasonics
- From:
Chinese Journal of Medical Instrumentation
2013;37(3):210-212
- CountryChina
- Language:Chinese
-
Abstract:
The cytotoxicity of medical ultrasonic couplant was tested by MTT assay and agar overlay test. By MTT assay, when the inoculum density was high, the cytotoxicity level was low, or vice versa. The cytotoxicity grade tested by agar overlay was not accord to MTT assay's too. MTT assay is suitable to test the cytotoxicity of medical ultrasonic couplant because it is quantitative and more sensitive, however, the experimental condition and the preparative method of extraction should be determined.