Comparison of ArcCheck and film verification for VMAT in the treatment of nasopharyngeal carcinoma.
- Author:
Jinling YI
1
;
Xiance JIN
;
Yongqiang ZHOU
;
Ce HAN
;
Xiaomin ZHENG
;
Zhiqin WU
;
Kejing HUANG
;
Congying XIE
Author Information
1. Department of Radiotherapy and Chemotherapy, 1st Hospital Affiliated to Wenzhou Medical College, Wenzhou 325000.
- Publication Type:Journal Article
- MeSH:
Carcinoma;
Humans;
Nasopharyngeal Neoplasms;
radiotherapy;
Radiotherapy Dosage;
Radiotherapy Planning, Computer-Assisted;
methods;
Software
- From:
Chinese Journal of Medical Instrumentation
2013;37(3):228-231
- CountryChina
- Language:Chinese
-
Abstract:
To compare the performance of ArcCheck and film verification for volumetric intensity modulated arc therapy (VMAT) in the treatment of nasopharyngeal carcinoma, and to study the feasibility of ArcCheck in VMAT dosimetric verification. Five patients of nasopharyngeal carcinoma treated with VMAT were enrolled in this study. Dose verification was carried out by ArcCheck and film respectively. The result showed that there were no significant differences between ArcCheck and film verification. ArcCheck software can obtain three dimensional dose distribution directly with simple operation. It is convenient for ArcCheck to be used for VMAT dosimetric verification.