Establishment and evaluation of intraocular lens loop support force test platform.
- Author:
Li LIU
1
;
Linnan KE
;
Shuo WANG
;
Ying WANG
;
Haiping REN
Author Information
- Publication Type:Journal Article
- MeSH: Lenses, Intraocular; Prosthesis Design
- From: Chinese Journal of Medical Instrumentation 2013;37(4):291-293
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVEEstablish the test platform of the intraocular lens loop, and the platform was evaluated through the experiment.
METHODSThe intraocular lens loop test platform is made up with three models. The different intraocular lens haptics support force can be completed by replacing different sample holder model.
RESULTSThe standard deviation and the coefficient of variation were calculated through the result of the fifteen samples. The standard deviation was 0.04 mN, and the coefficient of variation was 0.66%. The two values were in the acceptable range.
CONCLUSIONSThe platform was so stabilizing that it could be used to test support force of IOL loop. The different shapes of IOL could be tested on the platform through the replacement of the holder model.