Study on the release of copper ion and corrosion for intrauterine contraceptive device with two kinds of Cu-containing stent.
- Author:
Fengping ZOU
;
Shuze HUANG
;
Rong JI
;
Biao DING
;
Tianping YAO
- Publication Type:Journal Article
- MeSH:
Copper;
chemistry;
Corrosion;
Intrauterine Devices;
Ions;
chemistry;
Stents
- From:
Chinese Journal of Medical Instrumentation
2014;38(3):222-224
- CountryChina
- Language:Chinese
-
Abstract:
Two kinds of Cu-containing intrauterine contraceptive device (Cu-IUD) has been evaluated through the study on the release of copper ion and corrosion of copper surface in artificially formulated fluid. The result showed that the copper ions were released profusely from the stainless steel stents Cu-IUD in the first week of soak (25-12 microg/d), and then it became slowly and steady. The release of the copper ions from the titanium nickel shape memory alloy stents Cu-IUD were always slowly and steady in two months (0.5-5 microg/d). It obviously shows that the corrosion on copper surface of stainless steel stents Cu-IUD were more serious than titanium nickel shape memory alloy stents Cu-IUD by using 200 times microscope testing before and after soaking in formulated fluid.