Establishment of RP-HPLC detection method of N-isopropyl oxamate in the serum of plateau pikas.
- Author:
Yang WANG
;
Lian WEI
;
Lin-na WEI
;
Xiao LI
;
Li-na XU
;
Deng-bang WEI
- Publication Type:Journal Article
- MeSH: Animals; Chromatography, High Pressure Liquid; methods; Lagomorpha; Male; Oxamic Acid; analogs & derivatives; blood
- From: Chinese Journal of Applied Physiology 2015;31(5):469-476
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo explore the intergrating of N-isopropyl oxamate and serum protein and establish a high performance liquid chromatography (HPLC) detection method of N-isopropyl oxamate (specific inhibitor of testis-specific lactate dehydrogenase (LDH-C4)) in the blood of plateau pikas.
METHODSTwenty highland pika 150-200 g, were randomly divided into two groups (n = 10): control group and inhibitor group. Different concentrations of N-isopropyl oxamate were added to examine the intergrating of N-isopropyl oxamate and serum protein. In order to determine its concentration in the pika blood accurately, we used the method of adding trypsin to incubate the serum first, followed by trichloroacetic acid treatment and detecting by HPLC. Results: When the concentrations of N-isopropyl oxamate in the pika serum were added to 0.05 mmol/L, 0.1 mmol/L, 1 mmol/L, 10 mmol/L, 16.7 mmol/L, 33.3 mmol/L and 100 mmol/L, the intergrating rates between N-isopropyl oxamate and plateau pika serum were 100%, 100%, 100%, 86.84%, 54.11%, 40.10% and 20.18%, respectively. The method established in this paper was good on recovery rates, precision and stability. A good linearity was obtained in the range of 0.0125-0.25 mmol/L. When the concentrations of N-isopropyl oxamate in the serum were added to 0. 15 mmol/L,0.3 mmol/L and 1 mmol/L, the recovery rates were 98.05%, 98.98% and 98.12%, respectively; the precision relative standard deviation( RSD) of concentrations were 1.17%, 0.92% and 0.83%, respectively; the stability relative standard deviation (RSD) of concentrations were 1.38%, 1.40% and 0.88%, respectively. The repeatability RSD of the method was 1.76%. Quantitative limit was 0.0125 mmol/L.
CONCLUSIONN-isopropyl oxamate has a strong affinity with plateau pika serum protein that can't be accurately determined with common HIPLC method. It can be accurately determined in the blood by adding trypsinto digest the serum protein first, followed by adding trichloroacetic acid to precipitate the protein.