Exploitation of a new noninvasive animal's blood pressure measuring apparatus and its function characterization.
- Author:
Pei-Ci LI
1
;
Bin XU
;
Gu-Ya KONG
;
Mao-Hui ZHI
;
Shu-Ren WANG
Author Information
- Publication Type:Journal Article
- MeSH: Animals; Blood Pressure Determination; instrumentation; methods; Equipment Design; Rabbits; Software; Sphygmomanometers
- From: Chinese Journal of Applied Physiology 2006;22(1):122-124
- CountryChina
- Language:Chinese
-
Abstract:
AIMTo produce a new, noninvasive, animal's blood pressure measuring apparatus.
METHODSBy applying the principle similar with the measurement of human blood pressure, we developed a software which knowledge property owned by this research group and solved the problems of exactly demarcating the measuring spot of diastolic pressure. The blood pressure of rabbits was measured by the novel noninvasive animal's blood pressure measuring apparatus and the classic surgical catheterizing method simultaneously.
RESULTSThe novel noninvasive animal's blood pressure measuring apparatus was successfully set up. The measured blood pressures by this apparatus are very similar (r > 0.9) with the values obtained from the classic, surgical catheterizing method, no matter the blood pressures are normal, high or low.
CONCLUSIONOur new apparatus can be a reliable method for noninvasive measurement of the blood pressure of rabbits and rats.