Tissue Doppler imaging evaluate the effect of optimal biventricular resynchronization for congestive heart disease in left ventricular synchrony and function.
- Author:
Qian YANG
1
;
Jian-an WANG
;
Liang DONG
;
Hong HE
;
Xia SHENG
;
Yong SUN
;
Chen-yang JIANG
;
Bin-quan ZHOU
Author Information
- Publication Type:Journal Article
- MeSH: Aged; Cardiac Pacing, Artificial; methods; Echocardiography, Doppler, Color; Female; Heart Failure; diagnostic imaging; physiopathology; therapy; Humans; Male; Middle Aged; Ventricular Function, Left
- From: Chinese Journal of Cardiology 2005;33(12):1109-1113
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo evaluate the immediate change of left ventricular systolic performance and asynchronization between simultaneous biventricular pacing and sequential biventricular pacing by tissue synchronization imaging (TSI) and tissue velocity imaging (TVI) in patients with congestive heart failure. The effect of sequential biventricular resynchronization therapy was also observed.
METHODSTen patients with dilated cardiomyopathy who received sequential biventricular resynchronization were enrolled. The TVI and TSI imagings were performed by GE vivid7 with M3S probe. The left ventricular ejection fraction (LVEF), stroke volume (SV), aortic velocity time integral (VTI), left ventricular end diastolic diameter (LVEDd), the standard deviation of the electro-mechanical delay (EMD-SD) of 6 segments and TSI index were measured before implanting of InSync 8042 and 1 month, 3 months, 6 months after implanting respectively.
RESULTSAfter 6 months of implanting, the LVEF, SV and VTI were obviously increased from (22.0 +/- 8.8)% to (38.0 +/- 9.9)%; (36.0 +/- 14.9) ml to (57.0 +/- 15.7) ml; (20.22 +/- 5.72) cm to (26.20 +/- 5.98) cm, P < 0.05, respectively, compared with the before of implanting. The LVEDd was decreased from (6.6 +/- 0.6) cm, to (6.0 +/- 0.9) cm, P < 0.05. The EMD-SD and TSI-index were declined gradually after implanting, which was more evident in the 6 months after implanting, from (83.07 +/- 46.99) ms to (22.37 +/- 16.38) ms; (2.20 +/- 0.36) to (1.50 +/- 0.43), P < 0.05, respectively, but the immediate EMD-SD did not change obviously between simultaneous biventricular pacing and sequential biventricular pacing, whereas, the TSI index and VTI were significantly improved from (1.87 +/- 0.31) to (1.71 +/- 0.29); (22.44 +/- 5.43) cm to (25.44 +/- 5.36) cm, P < 0.05, respectively, in the sequential biventricular pacing.
CONCLUSIONSequential biventricular resynchronization could improve the left ventricular systolic function and synchronism of wall motion in the patients with congestive heart failure, which is more effective than simultaneous biventricular pacing after implanting immediately.