Establishment and evaluation of the diagnostic kit for anti-HIV1/2 antibody and P24 antigen.
- Author:
Yan HU
1
;
Jun HOU
;
Yan-qing FENG
;
Chang-fang FENG
;
Su-juan SHI
;
Hong-hui H SHEN
;
Zhi-jie WANG
;
Bao-jun WANG
;
Pan-yong MAO
Author Information
- Publication Type:Journal Article
- MeSH: Enzyme-Linked Immunosorbent Assay; HIV Antibodies; blood; HIV Core Protein p24; blood; HIV-1; immunology; HIV-2; immunology; Humans; Reagent Kits, Diagnostic
- From: Chinese Journal of Experimental and Clinical Virology 2007;21(4):391-393
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo establish and evaluate an Enzyme Immunoassay diagnostic kit combined with anti-HIV1/2 antibody and P24 antigen for shortening the examination window period of HIV infection in HIV laboratory diagnosis.
METHODSThe enzyme-linked reaction plates was coated by anti-HIV P24 monoclonal antibody and HIV 1/2 antigen. Labeling HIV1/2 antigen and anti-HIV P24 polyclonal antibody with horseradish peroxidase, setup an integrated ELISA kit for detecting anti-HIV-1/2 antibody and HIV P24 antigen, and evaluate the specificity and sensitivity of this kit.
RESULTSThe sensitivity of testing P24 antigen was up to 0.2 ng/ml. 78 serum samples of patients with AIDS, 85 serum samples of healthy people were compared with Abbott EIA kit, the coincidence was 100%. 12 051 sera from normal persons and patients were examined, the sensitivity of 100 %and specificity of 99.62 %, respectively.
CONCLUSIONThe anti-HIV1/2 antibody and HIV P24 antigen can be measured at the same time using this EIA kit, while the examination window period of HIV infection is shortened. Thus, the method is suitable for laboratory diagnosis and epidemiological investigation.