Comparative study of optic disc measurement by Copernicus optical coherence tomography and Heidelberg retinal tomography.
- Author:
Qing-Song YANG
1
;
Ya-Jie YU
;
Shu-Ning LI
;
Juan LIU
;
Ying-Juan HAO
Author Information
- Publication Type:Journal Article
- MeSH: Adolescent; Adult; Female; Humans; Male; Middle Aged; Optic Disk; pathology; Retina; pathology; Tomography; methods; Tomography, Optical Coherence; methods; Young Adult
- From: Chinese Medical Journal 2012;125(16):2858-2861
- CountryChina
- Language:English
-
Abstract:
BACKGROUNDCopernicus optical coherence tomography (SOCT) is a new, ultra high-speed and high-resolution instrument available for clinical evaluation of optic nerve. The purpose of the study was to compare the agreements between SOCT and Heidelberg retinal tomography (HRT).
METHODSA total of 44 healthy normal volunteers were recruited in this study. One eye in each subject was selected randomly. Agreement between SOCT and HRT-3 in measuring optic disc area was assessed using Bland-Altman plots. Relationships between measurements of optic nerve head parameter obtained by SOCT and HRT-3 were assessed by Pearson correlation.
RESULTSThere was no significant difference in the average cup area (0.306 vs. 0.355 mm, P = 0.766), cup volume (0.158 vs. 0.130 mm, P = 0.106) and cup/disc ration (0.394 vs. 0.349 mm, P = 0.576) measured by the two instruments. However, other optic disc parameters from SOCT were significantly lower compared with HRT-3. The Bland-Altman plot revealed good agreement of cup area and cup volume measured by SOCT and HRT-3. Bad agreement of disc area, rim area, rim volume and cup/disc ratio were found between SOCT and HRT-3. The highest correlations between the two instruments were observed for cup area (r(2) = 0.783, P = 0.000) and cup/disc ratio (r(2) = 0.669, P = 0.000), whereas the lowest correlation was observed for disc area (r(2) = 0.100, P = 0.037), rim area (r(2) = 0.275, P = 0.000), cup volume (r(2) = 0.005, P = 0.391) and rim volume (r(2) = 0.021, P = 0.346).
CONCLUSIONSThere were poor agreements between SOCT and HRT-3 for measurement of optic nerve parameters except cup area and cup volume. Measurement results of the two instruments are not interchangeable.