The development of a new perimembranous ventricular septal defect occluder.
- Author:
Zhi-wei ZHANG
1
;
Guo-hong ZENG
;
Shu-guang LIN
;
Rui-xin FAN
;
Yu-fen LI
;
Shu-shui WANG
;
Yu-mei XIE
;
Ji-jun SHI
;
Jun-jie LI
Author Information
- Publication Type:Journal Article
- MeSH: Adolescent; Adult; Animals; Balloon Occlusion; instrumentation; methods; Cardiac Catheterization; methods; Child; Child, Preschool; Equipment Design; Female; Heart Septal Defects, Ventricular; surgery; Humans; Male; Prosthesis Implantation; Swine; Treatment Outcome; Young Adult
- From: Chinese Journal of Cardiology 2005;33(3):228-231
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVEThe aim of this study was to develop a new perimembranous VSD occluder and to evaluate it.
METHODSThe shape of VSD occluder was designed as fabric frame "I" shape that comprised two types: symmetric and asymmetric. The safety, efficacy, feasibility and complication were tested in 22 animal models and in 58 VSD patients in clinical trial. The device were compared with Amplatzer occluder in this study.
RESULTSThe new perimembranous VSD occluder was passed the national material test. In animal study, artificial VSD were all occluded by using the new devices with no complication in follow up except one pig expresented wound infection. In clinical trial, all 58 VSD cases were healing with the new device. One patient suffered with atria-ventricular block 5 days after procedure and was free from AV block with medicine therapy. Compared with Amplatzer perimembranous VSD occluder, the new devices had lower frequency of residual shunt.
CONCLUSIONThe new perimembranous VSD occluder is a safe and effective perimembranous VSD interventional apparatus, and the effect of the new occluders seems not worse than that of the Amplatzer ones.