Differential scanning calorimetry analyses of phase transformations in different nickel-titanium orthodontic wires.
- Author:
Yu-xing BAI
1
;
Chao-chao REN
;
Hong-mei WANG
;
Yu-feng ZHENG
;
Yan-bo WANG
;
Song LI
Author Information
- Publication Type:Journal Article
- MeSH: Calorimetry, Differential Scanning; Dental Alloys; chemistry; Materials Testing; Orthodontic Wires; Phase Transition; Temperature
- From: Chinese Journal of Stomatology 2008;43(5):302-305
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo characterize austenite, martensite and R phase temperatures as well as transition temperature ranges of the commonly used nickel-titanium (NiTi) orthodontic arch wires selected from several manufacturers.
METHODSDifferential scanning calorimetry (DSC) method was used to study the phase transformation temperatures and the phase transition processes of 9 commonly used NiTi alloys (types: 0.406 mm, 0.406 mm x 0.559 mm).
RESULTSThe austenite finish temperatures of A, B, D NiTi wires were 22.4 CT, 21.9 degrees C, 22.5 degrees C, respectively. No phase transformation was detected during oral temperature. It indicated that these types of NiTi wires did not possess shape memory property. For C and H NiTi wires, no phase transformation was detected during the scanning temperature range, suggesting that these two types of wires did not possess shape memory either. The austenite finish temperatures of E, G and I NiTi wires were 34.3 degrees C, 36.6 degrees C, 38.5 degrees C, respectively, which were close to the oral temperature and presented as martensitic-austenitic structures at room temperature, suggesting that the NiTi wires listed above had good shape memory effect. Although F NiTi wire also showed martensitic-austenitic structures at room temperature, its austenite finish temperature (61.5 degrees C) was much higher than oral temperature.
CONCLUSIONSThe transformation phase temperatures and transformation behavior were varied among different NiTi alloys, leading to variability in shape memory effect.