Three-dimension finite element analyses of interior stress of two kinds of Replace implant.
- Author:
Jing LAN
1
;
Xin XU
;
Guangshui JIANG
;
Yanjin GUAN
;
Haiyun HUANG
;
Jin LIANG
;
Xiaoni MA
Author Information
- Publication Type:Journal Article
- MeSH: Alveolar Process; Dental Implants; Dental Stress Analysis; Finite Element Analysis; Humans; Mandible; Stress, Mechanical
- From: West China Journal of Stomatology 2011;29(5):464-468
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo establish three-dimension finite element model of mandible with two kinds of Replace implant, and to study stress of implant and abutment.
METHODSThe data of components of the dental implant was measured, cross section of the mandible was scanned by spiral CT and image reconstruction technique was conducted. Three-dimension finite element analysis software UG and MSC. Marc/Mentat were used to built the conjunction model and bone model of two implant systems. Axial loading (200N) and 30 degrees oblique loading (100N) were applied on the models respectively, the stress distribution patterns of the implant and abutment of two implant systems were analyzed.
RESULTSThe peak stress district was concentrated on the structure of the implant cervix, which was more obviously displayed on the Replace Select implant. The peak stress of oblique loading was higher than that of axial loading. The peak stress on the implant cervix of Replace Select implant was higher than that of Replace External Hex implant in all loadings.
CONCLUSIONTo Replace Select especially, oblique force should be avoided on clinical practice in case of the implant fracture.