- Author:
Ting WANG
1
;
Jie ZHANG
1
;
Juan WANG
1
;
Ying-Hua PEI
1
;
Xiao-Jian QIU
1
;
Yu-Ling WANG
1
Author Information
- Publication Type:Journal Article
- MeSH: Animals; Bronchoscopy; Disease Models, Animal; Dogs; Drug-Eluting Stents; Granulation Tissue; drug effects; Microscopy, Electron, Scanning; Paclitaxel; therapeutic use; Trachea; pathology; Tracheal Stenosis; drug therapy; surgery
- From: Chinese Medical Journal 2016;129(22):2708-2713
- CountryChina
- Language:English
-
Abstract:
BACKGROUNDCurrently available silicone and metallic stents for tracheal stenosis are associated with many problems. Granulation proliferation is one of the main complications. The present study aimed to evaluate the efficacy of paclitaxel drug-eluting tracheal stent in reducing granulation tissue formation in a canine model, as well as the pharmacokinetic features and safety profiles of the coated drug.
METHODSEight beagles were randomly divided into a control group (bare-metal stent group, n = 4) and an experimental group (paclitaxel-eluting stent group, n = 4). The observation period was 5 months. One beagle in both groups was sacrificed at the end of the 1st and 3rd months, respectively. The last two beagles in both groups were sacrificed at the end of 5th month. The proliferation of granulation tissue and changes in tracheal mucosa were compared between the two groups. Blood routine and liver and kidney function were monitored to evaluate the safety of the paclitaxel-eluting stent. The elution method and high-performance liquid chromatography were used to characterize the rate of in vivo release of paclitaxel from the stent.
RESULTSCompared with the control group, the proliferation of granulation tissue in the experimental group was significantly reduced. The drug release of paclitaxel-eluting stent was the fastest in the 1st month after implantation (up to 70.9%). Then, the release slowed down gradually. By the 5th month, the release reached up to 98.5%. During the observation period, a high concentration of the drug in the trachea (in the stented and adjacent unstented areas) and lung tissue was not noted, and the blood test showed no side effect.