Reconfirmation of Preanalytical Variables and Establishment of Reference Intervals of Platelet Function Analyzer-100 Closure Times in Korean Adults.
10.3343/kjlm.2007.27.5.318
- Author:
Young Uk CHO
1
;
Hyun Sook CHI
;
Seongsoo JANG
;
Chan Jeoung PARK
Author Information
1. Department of Laboratory Medicine, University of Ulsan College of Medicine and Asan Medical Center, Seoul, Korea. hschi@amc.seoul.kr
- Publication Type:Original Article ; English Abstract
- Keywords:
PFA-100 closure time;
Korean adult;
Reference interval;
Preanalytical variables
- MeSH:
ABO Blood-Group System;
Adult;
Female;
Humans;
Korea;
Male;
Middle Aged;
Platelet Function Tests/*instrumentation/standards;
Reference Values;
Regression Analysis;
Time Factors
- From:The Korean Journal of Laboratory Medicine
2007;27(5):318-323
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
BACKGROUND: Platelet Function Analyzer-100 (PFA-100, Dade-Behring, Germany) is an instrument that simulates in vivo hemostatic plug formation under high shear flow by measuring the time required to occlude aperture. We performed this study to reconfirm preanalytical variables and establish the reference intervals of Korean adults. METHODS: A total of 120 healthy individuals were enrolled. Closure times (CT) with the collagen/ epinephrine (CEPI) and the collagen/ADP (CADP) cartridges were measured. RESULTS: The reference intervals by the central 95th percentile were 82-182 sec for CEPI-CT and 62-109 sec for CADP-CT. Females had significantly longer CEPI- and CADP-CT than males (P=0.034 and 0.022, respectively). Individuals over 40 yr showed shorter CEPI- and CADP-CT compared with younger ones (P=0.002 and 0.003, respectively). CEPI- and CADP-CT values measured in the afternoon were significantly longer than corresponding ones in the morning (P<0.0001 in both conditions). Group O blood groups were related to longer CEPI- and CADP-CT compared with non-O blood groups (P=0.0003 and <0.0001, respectively). CADP-CT was weakly correlated with hematocrit (r=-0.296, P=0.001), but not CEPI-CT. CONCLUSIONS: We reconfirmed the preanalytical variables and established the reference intervals of PFA-100 CT in Korean adults. It is recommended that reference interval of this test should be verified according to age, diurnal variation, and ABO blood groups for optimal utilization.