Determination of the Phantom Scatter Factor (Sp Factor) using a small Block in the Phantom.
- Author:
Byong Yong YI
1
;
Seok Min HONG
;
Jae Sung KIM
;
Eun Kyung CHOI
;
Hyesook CHANG
;
Myung Za LEE
;
Ha Chung CHUN
Author Information
1. Department of Therapeutic Radiology, Asan Medical Center, College of Medicine, University of Ulsan, Seoul, Korea.
- Publication Type:Original Article
- Keywords:
Scatter Factor;
Dosimetry
- MeSH:
Gamma Rays;
Hepatocyte Growth Factor*
- From:Journal of the Korean Society for Therapeutic Radiology
1992;10(1):121-123
- CountryRepublic of Korea
- Language:English
-
Abstract:
New measurement method for Sp factors (Phantom Scatter Factors) is presented. The theoretical development of the approach is disscused showing that Sp factors can be obtained from three measurements of ionnization in a blocked, reference field and open field. This method has been tested using 60Co gamma rays. The results were within 1% deviation between the theory and the experiment for the Sp facter. The new method does not need air measurement, and we could determine the Sp p factors with a small piece of block.