HbA1c: A Review of Analytical and Clinical Aspects.
10.3343/alm.2013.33.6.393
- Author:
Cas WEYKAMP
1
Author Information
1. MCA Laboratory, Queen Beatrix Hospital, Winterswijk, Netherlands. c.w.weykamp@skbwinterswijk.nl
- Publication Type:Review
- Keywords:
HbA1c;
Diabetes;
Standardization;
Analytical aspects;
Clinical aspects
- MeSH:
Diabetes Mellitus/blood/diagnosis;
Hemoglobin A, Glycosylated/*analysis/standards;
Humans;
Immunoassay/standards;
Point-of-Care Systems;
Quality Control
- From:Annals of Laboratory Medicine
2013;33(6):393-400
- CountryRepublic of Korea
- Language:English
-
Abstract:
After the relationship between glycemic control and the HbA1c concentration was demonstrated, many tests have been developed to determine the HbA1c concentration. The test results are standardized to the International Federation of Clinical Chemistry (IFCC) Reference Measurement Procedure (RMP) in harmony with the efforts of the National Glycohemoglobin Standardization Program (NGSP). The longitudinal use of the test requires strict quality management including accreditation of the laboratory, a dedicated internal control design, participation in an external quality assessment (EQA) program (proficiency test), and careful consideration of pre- and post-analytical aspects of the test. Performance goals for optimizing determination of the HbA1c concentration have been described. As an index of long-term glycemic control and a risk predictor, the HbA1c concentration is an indispensable part of routine management of diabetes. Because of the improving quality of the test, the HbA1c concentration is being increasingly applied in the diagnosis of diabetes. There are, however, concerns of this application in point-of-care settings. The HbA1c concentration is also used to achieve stringent control in pregnant diabetic patients. Strict standardization enables the definition of universal reference values and clinical decision limits. This review describes the present status of analytical and clinical aspects of determining the HbA1c concentration and highlights the challenges involved.