Effect of Transverse Magnetic Field on Dose Distribution of High Energy Electron Beam.
- Author:
Young Kee OH
1
;
Ki Hwan KIM
;
Kyo Chul SHIN
;
Jhin Kee KIM
;
Jeung Kee KIM
;
Dong Hyeok JEONG
;
Mun Jun CHO
;
Jun Sang KIM
;
Sun Min YOON
;
Sung Kyu KIM
Author Information
1. Department of Radiation Oncology, Eulji University Hospital, Daejeon, Korea.
- Publication Type:Original Article
- Keywords:
Dose modification;
Transverse magnetic field;
Electron beam
- MeSH:
Magnetic Fields*
- From:Korean Journal of Medical Physics
2007;18(4):209-213
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
In this work we have measured the dose distribution and the percent depth dose of 20 MeV electron beam using the X-OMAT films in order to verify the effects of transverse magnetic field on high energy elecrtron beam in a phantom. The result shows about 30% increase of the percent depth dose at 4.5 cm depth under the transverse magnetic field of 1.5 Tesla at 7.5 cm depth. We have verified that these were in an agreement with other theoretical results.