A Study on the Development of R-R Interval Analyzer using Microcomputer (1).
10.12701/yujm.1985.2.1.77
- Author:
Joon Ha LEE
;
Soo Bong CHOI
- Publication Type:Original Article
- MeSH:
Autonomic Nervous System;
Methods;
Microcomputers*;
Periodicity;
Respiration
- From:Yeungnam University Journal of Medicine
1985;2(1):77-80
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
The R-R interval analyzer was developed to measure the autonomic nervous system function using microcomputer. The system based on 8 bit microcomputer including bandpass filter, R-wave detector and clock generator in order to obtain the mean value, standard deviation, total time, CV value, maximum value and minimum value in the specific view point of R-R interval variation. The pattern of R-R interval change after resting, voluntary standing and deep breathing can be analyzed in normal subjects and diabetics with autonomic nervous dysfunction. The amplitude of the R-R interval variation showed sensitive pattern for normal subjects at resting, standing and deep breathing. On the contrary, the periodicities of amplitude for abnormal subjects with autonomic nervous dysfunction showed dull pattern. It was suggested that R-R interval analyzer is a good detection method for dysfunction of autonomic nervous system.