Two Cases of WIOL-CF(R) IOL Dislocation after Nd:YAG Laser Capsulotomy.
10.3341/jkos.2015.56.3.443
- Author:
Byung Jae KIM
1
;
Ji Hye KIM
;
Seong Jae KIM
;
Yong Seop HAN
;
Jong Moon PARK
;
In Young CHUNG
Author Information
1. Department of Ophthalmology, Gyeongsang National University School of Medicine, Jinju, Korea. in0chung@hanmail.net
- Publication Type:Case Report
- Keywords:
IOL dislocation;
Nd:YAG laser capsulotomy;
WIOL
- MeSH:
Aged;
Aluminum;
Dislocations*;
Female;
Humans;
Lenses, Intraocular;
Vitrectomy;
Yttrium
- From:Journal of the Korean Ophthalmological Society
2015;56(3):443-446
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
PURPOSE: To report 2 cases of WIOL-CF(R) intraocular lens (IOL) (Gelmed International, Kamenne Zehrovice, Czech Republic) dislocation after neodymium-doped yttrium aluminum garnet (Nd:YAG) laser capsulotomy. CASE SUMMARY: A 78-year-old female was referred for IOL dislocation. She was implanted with WIOL-CF(R) IOL 18 months prior. Two months after WIOL-CF(R) implantation, she received Nd:YAG laser capsulotomy at a local clinic. Pars plana vitrectomy and transscleral fixation of IOL were performed. CONCLUSIONS: This is the first report of WIOL-CF(R) IOL dislocation after Nd:YAG laser capsulotomy in unvitrectomized eyes. When Nd:YAG laser capsulotomy is performed after WIOL-CF(R) IOL implantation, IOL dislocation should be considered even in unvitrectomized eyes.