An Artifact Caused by Using the Inadequate Acquisition Parameter in Tc-99m ECD Brain Perfusion SPECT.
- Author:
Han Kyung SEO
1
;
Min Woo KIM
;
Seok Tae LIM
;
Myung Hee SOHN
Author Information
1. Department of Nuclear Medicine, Chonbuk National University Medical School, Chonju, Korea.
- Publication Type:Case Report
- Keywords:
Artifact;
SPECT;
Brain;
Tc-99m ECD
- MeSH:
Artifacts*;
Brain*;
Perfusion*;
Tomography, Emission-Computed, Single-Photon*
- From:Korean Journal of Nuclear Medicine
2004;38(6):540-542
- CountryRepublic of Korea
- Language:English
-
Abstract:
Although artifacts in tomographic imaging can arise from a number of sources, this case presents an artifact caused by using inadequate acquisition parameters which do not match the collimator with which a camera equipped in Tc-99m ECD brain SPECT. Whenever the collimator in tomographic studies is changed, we should take care to use suitable parameters of acquisition protocol according to the collimator with which a camera is fitted. Other ways to prevent an error of this type is to use a point source.