Visual Fields in Normal Korean Measured with Humphrey Field Analyzer.
- Author:
Seong Jun LIM
1
;
Joo Hwa LEE
Author Information
1. Department of Ophthalmology, Inje University Seoul Paik Hospital, Seoul, Korea.
- Publication Type:Original Article
- Keywords:
Humphrey Field Analyzer;
Normal subjects;
Global indices
- MeSH:
Visual Field Tests;
Visual Fields*
- From:Journal of the Korean Ophthalmological Society
1998;39(7):1499-1504
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
Central 30-2 threshold tests were performed in 124 eyes of 62 normal subjects with Humphrey Field Analyzer. The average value of foveal threshold, mean deviation(MD), pattern standard deviation(PSD), shor-term fluctuation(SF) and corrected pattern standard deviation(CPSD) were 34.3+/-2.6dB, -1.96+/-1.31dB, 2.25+/-0.88dB, 1.38+/-0.52dB and 1.41+/-1.10dB, respectively. The frequency of abnormal value on automated perimetry(p<5%) was 25.5% in foveal threshold and 19.1% in MD, which was higher than fundamental data of perimetry. As the age increased, foveal threshold decreased (r=-0.729), PSD and CPSD increased (r=0.422) and 0.385, respectively) ; these are statistically significant correlations(p<0.05). This study has the limitation that test was performed on single session and had small study group. Therefore, further study is needed with comparative analysis following repeated test and larger group of visual field examination.