Comparison of Electroretinogram Waveforms Acquired Using Monopolar ERG-Jet Electrode and Bipolar Burian-Allen Electrode.
10.3341/jkos.2010.51.9.1217
- Author:
Jihyun PARK
1
;
Hokyung LEE
;
Sung Joon PARK
;
Yoon Hwa KIM
Author Information
1. Department of Ophthalmology, Kim's Eye Hospital, Konyang University College of Medicine, Seoul, Korea. hokyung@kimeye.com
- Publication Type:Original Article
- Keywords:
Burian-Allen electrode;
Electroretinogram;
ERG-Jet electrode
- MeSH:
Electrodes;
Eye
- From:Journal of the Korean Ophthalmological Society
2010;51(9):1217-1223
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
PURPOSE: To compare electroretinogram (ERG) waveforms acquired using an ERG-jet electrode and a Burian-Allen electrode. METHODS: ERGs were recorded with an ERG-jet electrode and a Burian-Allen electrode from both eyes of 29 volunteers. Three consecutive recordings were obtained with both electrodes from the left eye of another 6 volunteers. Peak-to-trough amplitudes and peak implicit times were compared between both eyes, and between the 2 types of electrodes. Interpersonal and intrapersonal variation were also compared. RESULTS: The differences in amplitude and implicit time between the right and left eyes were not influenced by the type of electrode. The amplitude of the ERG acquired using the Burian-Allen electrode, however, showed significant differences between both eyes, contrary to the amplitude acquired using the ERG-jet electrode. ERG recordings obtained using the ERG-jet electrode generally showed higher amplitude, shorter implicit time, and less interpersonal and intrapersonal variation than ERG recordings obtained using the Burian-Allen electrode. CONCLUSIONS: In this application, the ERG-jet ERG electrode appeared to be superior to the Burian-Allen electrode in terms of both usefulness and consistency.