The Three Dimensional Finite Element Analysis of Stress according to Implant Thread Design under the Axial Load.
- Author:
Woo Taek KIM
1
;
Yong Doo CHA
;
Se Jong OH
;
Sang Soo PARK
;
Hyun Woo KIM
;
Yang Ho PARK
;
Jun Woo PARK
;
Gun Joo RHEE
Author Information
1. Department of Oral & Maxillofacial Surgery, College of Medicine, Hallym University.
- Publication Type:Original Article
- MeSH:
Finite Element Analysis*
- From:Journal of the Korean Association of Oral and Maxillofacial Surgeons
2001;27(2):111-117
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
There are three designs of thread form in screw type implants: V-thread, Reverse buttress thread and Square thread. The purpose of this study was to find out how thread form designs have an influence on the equivalent stress, equivalent strain, maximum shear stress and maximum shear strain and which design of thread form generates more maximum equivalent stress and strain. 3-D finite element analysis was used to evaluate the stress and strain patterns of three tread types. The results of this study were as follow. 1. Under the 200N of axial load, the value of maximum equivalent stress is smallest in square thread and there is no significant difference between that of V thread and reverse buttress thread. 2. Under the 200N of axial load, the value of maximum equivalent strain is largest in V thread and smallest in square thread. 3. Under the 200N of axial load, the value of maximum shear stress is smallest in square thread and there is no significant difference between that of V thread and reverse buttress thread. 4. Under the 200N of axial load, the value of maximum equivalent strain is largest in V thread and there is no significant difference between that of square thread and reverse buttress thread. 5. Above results show that the square thread has special advantages in stress and strain compared with other thread types, especially in shear stess which is most determinant to implant-bone interface. Considering the superior biomechanical properties of square form implant, we presume that square form implant has better clinical results than the other types of implants in the same clinical conditions.