Performance Evaluation of the VISTA 500.
10.15263/jlmqa.2016.38.1.52
- Author:
Dae Hyun KO
1
;
Eun Jung CHO
;
Woochang LEE
;
Sail CHUN
;
Won Ki MIN
Author Information
1. Department of Laboratory Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea. wkmin@amc.seoul.kr
- Publication Type:Brief Communication
- Keywords:
VISTA 500;
Performance;
Evaluation studies as topic
- MeSH:
Allergy and Immunology;
Ceruloplasmin;
Cystatin C;
Evaluation Studies as Topic;
Haptoglobins;
Humans;
Immunoassay;
Immunoglobulin A;
Immunoglobulin G;
Immunoglobulin M;
Prealbumin;
Rheumatoid Factor;
Statistics as Topic;
Transferrin
- From:Journal of Laboratory Medicine and Quality Assurance
2016;38(1):52-57
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
In this study, we evaluated the performance of a recently developed immunoassay analyser, the VISTA 500 (Siemens, Germany). Precision, linearity, and comparison studies were performed according to the Clinical and Laboratory Standards Institute guidelines. The test items evaluated included IgG, IgA, IgM, C3, C4, ceruloplasmin, prealbumin, transferrin, haptoglobin, rheumatoid factor, anti-streptolysin O, and cystatin C. Commercial control materials (BioRad Laboratories, USA), commercial linearity validation materials (Maine Standards, USA), and patient samples were used for the evaluation. For the correlation study, analysis with a BN-II nephelometer (Siemens) was used as a comparative method. Total coefficients of variation of analytes were found to be between 1.9% and 5.5%. Results of the linearity evaluation were also acceptable for the range tested. Correlations with comparative methods were acceptable. The VISTA 500 analyser showed satisfactory analytical performance with respect to precision, linearity, and comparison. We conclude that the VISTA 500 is likely a good candidate as an immunology analyser.