Comparison of Two Commercial Real-time Nucleic Acid Amplification Methods for Detecting the Viral Load of Human Immunodeficiency Virus Type 1 .
- Author:
Sue Jung KIM
1
;
Eunkyung LEE
;
Younhee PARK
;
Hyon Suk KIM
Author Information
1. Department of Laboratory Medicine, Yonsei University College of Medicine, Seoul, Korea. kimhs54@yuhs.ac
- Publication Type:Original Article
- Keywords:
HIV;
Viral load;
RT-PCR;
NASBA
- MeSH:
Decision Making;
HIV;
HIV-1;
Humans;
Limit of Detection;
Prognosis;
RNA;
Self-Sustained Sequence Replication;
Viral Load
- From:Korean Journal of Blood Transfusion
2008;19(3):216-221
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
BACKGROUND: Detection of the viral load of Human Immunodeficiency Virus type 1 (HIV-1) RNA is important for clinical decision making and for determining the prognosis of HIV-infected patients. The aim of the study is to compare the performance of real-time RT-PCR (COBAS AmpliPrep/COBAS TaqMan HIV-1, CAP/CTM, Roche Diagnostics) and the Nucleic Acid Sequence Based Amplification (NucliSens EasyQ HIV-1, NucliSens, BioMerieux) methods for testing Korean HIV-infected patients. METHODS: Among the specimens that were requested to undergo HIV-1 RNA viral load detection from 2005 to 2006, 153 specimens were selected based on the status of the specimens. The CAP/CTM and NucliSens tests were performed according to the manufacturer's instruction. RESULTS: HIV-1 RNA was detected by both tests in 93 specimens. Among the remainder, CAP/CTM detected HIV-1 RNA in 10 specimens, while the same specimens showed results lower than the detection limit with using the NucliSens. Though the results were appropriately correlated (r=0.85, P<0.0001), the mean differences between the two test results were -0.1321 log(10) IU/mL on the Bland-Altman test. CONCLUSION: The methodologic difference or the presence of a HIV subtype may affect the agreement between the two tests. The standardization of methods and the establishment of a linear range for the individual laboratory results may be helpful to obtain accurate test results.