Analytical Evaluation of the DiaSys Albumin in Urine/CSF FS Kit for Urine Albumin Measurement Using a JEOL BioMajesty JCA-BM6010/C Analyzer.
- Author:
Hyun Jeong KIM
1
;
Misuk JI
;
Hee Won MOON
;
Mina HUR
;
Yeo Min YUN
Author Information
- Publication Type:Original Article
- Keywords: High albuminuria; Albumin-to-creatinine ratio; Performance
- MeSH: Albuminuria
- From:Laboratory Medicine Online 2016;6(2):64-69
- CountryRepublic of Korea
- Language:Korean
- Abstract: BACKGROUND: High albuminuria is defined as albumin excretion of >30 mg/24 hr or an albumin-to-creatinine ratio of 30 mg/g in a random urine sample. We assessed the analytical performance of the Albumin in Urine/CSF FS kit (DiaSys Inc., UK) using a BioMajesty JCA-6010/C analyzer (JEOL Inc., Japan). METHODS: Urine albumin concentrations were measured by the Albumin in Urine/CSF FS kit using a BioMajesty JCA-BM6010/C analyzer. Imprecision, linearity, and carry-over were measured according to the Clinical Laboratory and Standards Institute documents EP10 and EP9. The assay was compared with the ALB-T TQ Gen.2 (Roche, Germany) assay on a Cobas8000 C702 (Roche, Germany), the Tina-Quant Albumin (Roche, Switzerland) assay on a Hitachi7600-210 (Hitachi, Japan), and an Abbott urine albumin assay (Abbott Laboratories, USA) on a TBA 200FR (Toshiba, Japan) using 50 random urine samples. RESULTS: Within-run and total imprecision were 0.551-1.023% and 0.551-1.214%, respectively. Linearity ranged from 6.31 to 30.60 mg/dL, and functional sensitivity was 0.5 mg/dL. Results from the Albumin in Urine/CSF FS kit showed good correlation with the ALB-T TQ Gen.2 (r=0.987) and the Tina-Quant Albumin assays (r=0.991). However, the four assays categorized 18 of 50 urine samples into different albuminuria groups. CONCLUSIONS: Albumin in Urine/CSF FS testing on a BioMajesty JCA-BM6010/C analyzer showed good linearity, functional sensitivity, precision, and correlation with the ALB-T TQ Gen.2 and Tina-Quant Albumin assays. However, because some samples were categorized into different albuminuria groups by the different assays, further studies on the standardization of albuminuria assays are needed.